Key facts
- Frequency range upgradeable to 10 MHz
- Basic accuracy of ±0.05 % for impedance measurements
- Basic accuracy of ±0.03° for phase measurements
- DC bias up to 40 V
- Data logging function
The top-class in component testing
The R&S®LCX LCR meters are versatile, extremely accurate and perform measurements quickly. They are ideal for challenging applications in research, development and production. Two instrument models and various options cover applications with a test signal frequency of up to 10 MHz. Internal and external bias functions, comprehensive analysis options and versatile test fixtures expand the wide range of possible applications.
Fast, accurate and versatile
Both R&S®LCX models combine high measurement speed, accuracy and versatile measurement capabilities. This makes them ideal for standard measurements in development, for material analysis in research as well as for fast production testing. With their broad measurement ranges, they also cover applications with extremely low and extremely high impedances. Three measurement times are available:
- Fast: ≤ 15 ms
- Medium: ≤ 100 ms
- Slow: ≤ 500 ms
The basic accuracy for impedance measurements is ±0.05 %, and for phase measurements ±0.03°.
Selectable frequency range
All R&S®LCX models measure under DC conditions. The AC range already begins at 4 Hz. The upper frequency limit on the R&S®LCX100 is 300 kHz. In its base configuration, the R&S®LCX200 is designed for a maximum frequency of 500 kHz; this frequency limit can, however, be extended to 1 MHz or 10 MHz. This means that the ideal instrument is available for any given application and budget.
Test signals for all requirements
Test signals can be generated from 100 mV to 10 V and deliver a current of up to 200 mA. The instruments have a selectable output impedance of 100 Ω or 10 Ω. The actual current flow and the voltage applied are measured using the monitor function.
DC bias
In many applications, an adjustable DC bias is needed to measure C and L components at different operating points. The R&S®LCX100 and R&S®LCX200 generate a DC bias voltage of up to 10 V. As an option, it is also possible to set a DC bias current (up to 200 mA). DC bias voltages of up to 40 V can be applied at an external connection on the rear panel (R&S®LCX-K108 option) using a standard DC power supply, for example an R&S®NGA.
Data logging function
The R&S®LCX LCR meters provide a fast logging function for recording all measured values. The data can be saved on an external USB flash drive or transferred to an external PC via USB or LAN. With a data rate of up to 10 sample/s, the measured values are available every 100 ms.
Model | Description |
---|---|
R&S®LCX100 | LCR meter, meas. frequency up to 300 kHz test signals up to 10 V / 200 mA, bias: current intern, voltage int./ext., RDC and transformer measurements, USB/LAN interface |
R&S®LCX200 | LCR meter, meas. frequency up to 500 kHz optional up to 10 MHz, test signals up to 10 V / 200 mA, bias: current intern, voltage int./ext., RDC and transformer measurements, USB/LAN interface |
R&S®LCX200 Complete Package |
Package LCX200 COMPLETE PACKAGE Contains serialized product+options: R&S®LCX200 LCR METER ADVANCED 3629.8856.03 Consisting of: -R&S®LCX200 LCR METER ADVANCED -R&S®LCX-K106 ADVANCED ANALYSIS OPTION -R&S®LCX-K107 BINNING OPTION -R&S®LCX-K108 EXTENDED BIAS OPTION -R&S®LCX-K210 FREQUENCY UPGRADE 10 MHZ |